科研 Skill

电刺激安全参数核查

neuro-stim-safety-check

Checks electrical stimulation parameters against charge-density and electrode-material limits such as the Shannon criterion. Use when reviewing stimulation amplitude or pulse width.

发布者 Codar安全审核通过v0.1.0更新于 2026-11-18

SKILL.md

Stimulation safety check

Scope: research engineering support. Final settings for humans or animals follow the approved protocol, device labeling and the responsible clinician or ethics/IACUC committee.

Two different limits apply and both must hold:

  1. Tissue damage, usually screened with the Shannon criterion for macroelectrodes.
  2. Electrode electrochemistry: the charge the electrode material can inject reversibly without gas evolution or dissolution.

1. Compute with the bundled calculator

python scripts/stim_calc.py --current-ua 60 --pw-us 200 --freq-hz 100 \
       --area-um2 2000 --material platinum [--shannon-k-limit 1.85]

Geometry options: --area-cm2, --area-mm2, --area-um2, --disk-diam-mm, or --cylinder-diam-mm D --cylinder-len-mm L (lateral area, as for DBS ring contacts). Voltage-mode stimulators: --voltage-v V --impedance-ohm Z (current estimated as V/Z; impedance changes with amplitude and over time, so treat as approximate). Asymmetric waveforms: --anodic-ua and --anodic-pw-us for the second phase.

It prints charge per phase, charge density, Shannon k, net charge per pulse and DC current, microelectrode warning, and the comparison with the material's approximate reversible charge-injection range.

2. Interpret

  • Shannon k (log D = k - log Q; D in uC/cm^2/phase, Q in uC/phase, Shannon 1992): k = 1.85 is the boundary in the original data; many groups use k = 1.5 as a conservative limit. It was derived from macroelectrode data; microelectrodes do not follow it (Cogan et al. 2016), so for small sites rely on electrode and microstimulation-specific data instead.
  • Material limit: approximate reversible ranges (Cogan 2008): platinum and platinum-iridium about 50-150 uC/cm^2, activated iridium oxide about 1-5 mC/cm^2, titanium nitride about 0.9 mC/cm^2 (all geometric area, waveform dependent). Above the range, measure voltage transients (see neuro-electrode-electrochemistry) before use.
  • Charge balance: net charge per pulse must be near zero; any residual is DC current into tissue. Check passive recharge and blocking capacitors.
  • Device labeling: clinical devices carry their own limits (for example 30 uC/cm^2 per phase in some DBS labeling). Check the actual labeling of the device in use.

3. Report

Inputs with units, computed Q, D, k, net DC, material comparison, which criteria apply and which do not, the verdict per criterion, and what to measure next. Start the report with the research-only scope line.

示例提示词

帮我检查这组DBS刺激参数的电荷密度是否超过安全限值。

所属插件

神经科学面向神经科学研究:EEG/MEG、fMRI、BIDS、spike sorting、连接组、脑机接口解码评估、神经刺激安全与公开神经数据检索。

安装该插件会一并获得同领域的 45 个 Skill / MCP。

同插件的其他科研 Skill

常见问题

电刺激安全参数核查能做什么?

Checks electrical stimulation parameters against charge-density and electrode-material limits such as the Shannon criterion. Use when reviewing stimulation amplitude or pulse width。

如何安装电刺激安全参数核查?

在科搭 Codar 左侧点击「插件」,在「组件」页签搜索「电刺激安全参数核查」并安装;在 Claude Code 或 Codex 中可以发送本页的安装提示词,或下载 Skill 包后解压到 ~/.claude/skills 或 ~/.codex/skills。

电刺激安全参数核查需要哪些权限?

该科研 Skill没有声明额外权限;执行具体操作时仍由 Agent 按自身规则请求确认。

电刺激安全参数核查经过官方审核吗?

是。SpeedSkills 收录的每个版本都会记录内容摘要与权限声明,当前版本已通过安全审核。