Characterizes neural electrodes electrochemically via impedance spectroscopy, cyclic voltammetry and charge-injection limits. Use when comparing electrode coatings or charge capacity.
SKILL.md
Electrode electrochemistry
Recording quality depends on impedance; stimulation capacity depends on how much charge the interface injects reversibly. Measure both with standard methods and report conditions, because values change with electrolyte, sweep rate, pulse width and bias.
0. Load the tool reference
Read references/instruments-and-fitting.md: Gamry, BioLogic, Autolab
and PalmSens exports, how to fit spectra (scripts/fit_eis.py recovered
known values where impedance.py's default fit did not), CSC and voltage
transient analysis.
1. Impedance spectroscopy
- Three-electrode cell (working, large counter, Ag/AgCl reference) in phosphate-buffered saline at room or body temperature; state which.
- Sweep about 1 Hz to 100 kHz with a small sinusoid (about 10 mV rms).
- Report |Z| and phase at 1 kHz (the field's common comparison point) plus the full spectrum.
- Fit an equivalent circuit to separate solution resistance, charge- transfer resistance and double-layer behaviour:
python scripts/fit_eis.py --data eis.csv [--model randles-cpe]
eis.csv columns: freq_hz,z_mag_ohm,phase_deg. The script fits
Rs + (Rct || CPE) and reports Rs, Rct, CPE Q and alpha, the 1 kHz
impedance, and fit residuals.
2. Cyclic voltammetry and CSC
Sweep within the water window of the material (for platinum commonly about -0.6 to +0.8 V vs Ag/AgCl) at a stated sweep rate. Cathodic CSC is the time integral of cathodic current over the cycle divided by geometric area. CSC at slow sweep greatly overestimates what a short pulse can deliver; do not use it as the charge-injection limit.
3. Voltage transients and charge-injection limit
Apply the actual stimulation pulse (current, pulse width, interphase, bias) and record electrode voltage. Subtract the access-voltage step (Va, the instantaneous ohmic drop) to get the maximum cathodic polarisation (Emc). The charge-injection limit is the charge density at which Emc reaches the water-reduction limit for the material. Repeat in vivo when possible; in vivo limits are usually lower than in saline (Cogan 2008).
4. Typical ranges (orientation only)
Approximate reversible charge-injection ranges (geometric area, Cogan 2008): platinum about 50-150 uC/cm^2, activated iridium oxide about 1-5 mC/cm^2, titanium nitride about 0.9 mC/cm^2; conductive polymers report higher values that depend strongly on deposition and stability. Use them to sanity-check, not to replace measurement.
Output
Setup and conditions, EIS at 1 kHz and fit parameters, CSC with sweep
rate, voltage-transient Emc and resulting charge-injection limit, and the
comparison with the intended stimulation (see neuro-stim-safety-check).
Limitations
The typical charge-injection ranges in step 4 are orientation values from
the literature (Cogan 2008), not a guarantee for a specific electrode,
coating batch or pulsing regime; in vivo values are usually lower than in
saline and can change with chronic use (corrosion, biofouling, coating
delamination). Measured limits inform research and device-design decisions
but are not a clinical or regulatory safety sign-off; final stimulation
settings for human or animal use require review under the applicable
protocol (see neuro-stim-safety-check and neuro-device-regulatory).
示例提示词
帮我分析这组PtIr电极的阻抗谱和电荷存储容量测试数据。
所属插件
安装该插件会一并获得同领域的 45 个 Skill / MCP。
同插件的其他科研 Skill
- Allen脑区本体查询Looks up Allen Institute brain region names, acronyms and atlas hierarchy, and picks a BrainGlobe reference atlas. Use when matching region labels across datasets or papers.
- 动物行为姿态追踪Tracks animal pose and behavior from video with DeepLabCut or SLEAP, and extracts kinematic and behavioral event features. Use for markerless tracking or scoring behavior.
- BIDS神经影像数据管理Converts raw EEG or fMRI data to BIDS, validates with bids-validator, and queries or manages datasets with pybids and datalad. Use when organizing or sharing neuroimaging data.
- Brian2脉冲神经网络仿真Simulates LIF, AdEx and Hodgkin-Huxley spiking neuron networks in Brian2, including STDP plasticity and recurrent dynamics. Use when building a spiking network model.
- 钙成像数据分析Processes two-photon or widefield calcium imaging with Suite2p or CaImAn, from motion correction through ROI extraction to dF/F and spike deconvolution. Use for calcium imaging movies.
- 功能连接组图论分析Computes functional connectivity matrices and graph-theory metrics such as modularity and path length from ROI time series. Use when analyzing brain network topology.
- MNE脑电脑磁分析Runs an MNE-Python EEG or MEG pipeline covering preprocessing, ICA cleanup, ERP and time-frequency analysis, and source localization. Use to analyze raw EEG/MEG recordings end to end.
- 神经解码无泄漏评估Guides leakage-free cross-validation of neural decoders, including subject grouping, nested CV, chance levels and confound checks. Use when training or reporting a BCI classifier.
- 电极通道映射校验Checks that EEG, ECoG or probe channel labels and positions match physiology using correlation-distance consistency. Use when montage labels may be swapped or bridged.
- 聚类置换统计检验Explains how to set up and correctly interpret cluster-based permutation tests and TFCE for ERP or time-frequency data. Use to avoid onset or localization overclaims.
- 神经连接性分析核查Guides choice of leakage-robust connectivity measures such as wPLI, imaginary coherency and AEC, with confound controls. Use when reviewing or planning a connectivity analysis.
- 神经数据集陷阱审查Checklist for auditing neural datasets before training: rate and channel consistency, duplicate subjects, label mapping. Use before modeling on a public or lab dataset.
- 神经数据档案库导航Lists neural data archives such as OpenNeuro, DANDI, PhysioNet and MOABB by paradigm, with access tiers and download guidance. Use when finding a dataset for a research question.
- DBS植入物传感数据分析Loads, time-aligns and analyzes sensing data from DBS and bidirectional implants such as Percept and Summit RC+S. Use with implant LFP exports, packet loss or aDBS biomarkers.
- 神经解码器嵌入式移植Ports a neural signal-processing or BCI decoder from Python or MATLAB to fixed-point embedded C on a microcontroller. Use when moving a decoder onto implant or headstage firmware.
- 神经设备法规合规映射Maps a neurotech device to the regulatory standards and testing it needs and builds a requirements-to-risk traceability matrix. Use when planning verification or a design review.
- EEG预处理流程审查Designs or audits an EEG or MEG preprocessing pipeline in MNE-Python, EEGLAB, FieldTrip or Brainstorm. Use when choosing or reviewing filter, ICA or epoching settings.
- 神经记录文件格式转换Opens, inspects and converts neural recording files from EEG, MEG, intracranial and fNIRS acquisition systems. Use when a reader errors or units, rates or events look wrong.
- 神经信号基础模型评估Fairly evaluates pretrained EEG or iEEG foundation models against classical baselines with matched data and splits. Use when fine-tuning or linear-probing a pretrained model.
- 神经记录前端电路设计Designs or reviews analog front ends for neural recording headstages, covering noise budget, CMRR and ADC resolution. Use when choosing an amplifier or debugging a noisy front end.
- 脑机解码器长期稳定性Evaluates intracortical BCI decoders across days using zero-shot and few-shot regimes to quantify drift and recalibration cost. Use to assess decoder stability over sessions.
- 植入设备功耗热预算Budgets power, data rate and tissue heating for implanted or wireless neural interfaces against thermal screening limits. Use when sizing a high-channel-count implant design.
- NeuroKit2生理信号处理Processes physiological biosignals such as ECG, EEG, EDA, RSP, EMG and EOG with NeuroKit2, including HRV and complexity metrics. Use for psychophysiology or multi-modal signal analysis.
- NeuroMorpho神经元形态检索Finds and compares digital neuron reconstructions on NeuroMorpho.Org by region, cell type or species and reads their morphometry. Use when real neuron morphologies are needed.
常见问题
神经电极电化学表征能做什么?
Characterizes neural electrodes electrochemically via impedance spectroscopy, cyclic voltammetry and charge-injection limits. Use when comparing electrode coatings or charge capacity。
如何安装神经电极电化学表征?
在科搭 Codar 左侧点击「插件」,在「组件」页签搜索「神经电极电化学表征」并安装;在 Claude Code 或 Codex 中可以发送本页的安装提示词,或下载 Skill 包后解压到 ~/.claude/skills 或 ~/.codex/skills。
神经电极电化学表征需要哪些权限?
该科研 Skill没有声明额外权限;执行具体操作时仍由 Agent 按自身规则请求确认。
神经电极电化学表征经过官方审核吗?
是。SpeedSkills 收录的每个版本都会记录内容摘要与权限声明,当前版本已通过安全审核。