Checks that EEG, ECoG or probe channel labels and positions match physiology using correlation-distance consistency. Use when montage labels may be swapped or bridged.

发布者 Codar安全审核通过v0.1.0更新于 2026-11-18

SKILL.md

Channel-map verification

A wrong map produces plausible-looking but wrong topographies, source estimates and spike locations. Check the map against the physics of the signal: nearby electrodes look alike, and the pattern of similarity should follow the geometry.

1. Similarity should follow distance

Compute the correlation matrix of a few minutes of broadband or alpha-band data and compare with the distance matrix from the montage or probe file. Neighbouring sites should correlate most. For each channel, list its five most correlated channels and check they are physical neighbours. Channels whose best partners are far away are mislabelled or swapped; blocks of channels behaving this way point to a swapped connector or headstage (for example a 32-channel block offset).

2. Bridged electrodes (EEG)

Gel bridges make two electrodes nearly identical. Compute the variance of the difference between each pair (electrical distance; Tenke & Kayser 2001; Alschuler et al. 2014). Pairs with near-zero difference variance compared with other neighbours are bridged. Re-prep or treat the pair as one channel; do not interpolate one from the other.

3. Physiological landmarks

  • EEG: alpha largest at occipital sites with eyes closed; blinks largest at Fp1/Fp2 with frontal-to-posterior decay; lateral eye movements reverse polarity between F7 and F8. Wrong patterns mean a wrong montage.
  • ECoG/sEEG: line noise and artifacts shared within a shaft or grid; check contact order along a shaft from the depth gradient of signals.
  • Laminar or Neuropixels probes: LFP current-source density should show a plausible laminar profile; a spike's waveform should spread over a few adjacent sites in the probe geometry. Spikes "jumping" between distant sites indicate a wrong geometry file.
  • Utah arrays: compare the map from the manufacturer's connector file; correlations should form a grid, not random pairs.

4. Fix and record

Correct the mapping file, not the data; keep the old and new map with the evidence (correlation-vs-distance plot). Re-run downstream analyses that used positions.

Output

Map checked, method, channels or blocks found swapped, bridged pairs, evidence plot description, corrected mapping.

示例提示词

帮我检查这份EEG montage的电极通道标签有没有接反或错位。

所属插件

神经科学面向神经科学研究:EEG/MEG、fMRI、BIDS、spike sorting、连接组、脑机接口解码评估、神经刺激安全与公开神经数据检索。

安装该插件会一并获得同领域的 45 个 Skill / MCP。

同插件的其他科研 Skill

常见问题

电极通道映射校验能做什么?

Checks that EEG, ECoG or probe channel labels and positions match physiology using correlation-distance consistency. Use when montage labels may be swapped or bridged。

如何安装电极通道映射校验?

在科搭 Codar 左侧点击「插件」,在「组件」页签搜索「电极通道映射校验」并安装;在 Claude Code 或 Codex 中可以发送本页的安装提示词,或下载 Skill 包后解压到 ~/.claude/skills 或 ~/.codex/skills。

电极通道映射校验需要哪些权限?

该科研 Skill没有声明额外权限;执行具体操作时仍由 Agent 按自身规则请求确认。

电极通道映射校验经过官方审核吗?

是。SpeedSkills 收录的每个版本都会记录内容摘要与权限声明,当前版本已通过安全审核。